Supercritical Fluid Solubility Measurement System

Model: HPA-SM

The proposed FLOSN HPA-SM uses a dynamic saturation method for compatible low-volatility solutes in supercritical CO₂. A cooled feed pump, serial saturators, temperature-controlled pressure reduction and tandem collection traps support timed sampling with solvent totalization. Collected solute is quantified by a separately validated gravimetric or analytical method. The platform is designed for equilibrium-solubility method development, not a universal solubility detector.

FLOSN / PRODUCT FEATURES

Product Features

6 FEATURES

Serial Saturation Path

Two cartridges provide staged contact and a route for saturation-method checks.

Defined Collection Intervals

Solute collection and solvent totalization share the same accepted time window.

Heated Pressure Reduction

Transfer and back-pressure components are temperature controlled to manage deposition risk.

Tandem Recovery Checks

The second trap provides evidence for assessing breakthrough from the first.

Method-Specific Quantification

Validated weighing, chromatography or spectroscopy can quantify recovered solute.

Transparent Mass-Balance Basis

Reports distinguish recovered mass, CO₂ quantity, concentration basis and recovery corrections.

FLOSN / TECHNICAL PARAMETERS

Product Features

SELECTED MODEL

CFD-SA1000

Model-specific specifications
Chamber volume
1,080 mL
Operating pressure
0–2,200 psi
Power consumption
1,500 W
Exhaust flow rate
0–2 SLPM

Chamber & operation

Chamber dimensions · I.D. × depth (mm)
165.1 × 50.8 mm
Chamber dimensions · I.D. × depth (in)
6.5 × 2 in
Operating pressure · MPa
0–15.2 MPa
Maximum safe working pressure
2,500 psi (17.2 MPa)
Heating method
Cartridge heater heating
Overall dimensions · W × D × H
Not specified
SELECTED MODEL

CFD-SA1800

Model-specific specifications
Chamber volume
1,860 mL
Operating pressure
0–2,200 psi
Power consumption
2,000 W
Exhaust flow rate
0–0.5 SLPM

Chamber & operation

Chamber dimensions · I.D. × depth (mm)
215.9 × 50.8 mm
Chamber dimensions · I.D. × depth (in)
8.5 × 2 in
Operating pressure · MPa
0–15.2 MPa
Maximum safe working pressure
2,500 psi (17.2 MPa)
Heating method
Cartridge heater heating
Overall dimensions · W × D × H
Not specified
Shared specifications · all models

Materials & pressure safety

Chamber material
SS316L
Optional materials
Hastelloy C276, Inconel 625, Monel 400, Titanium Grade 2, PTFE lining
Pressure safety system
Electronic pressure monitoring combined with mechanical over-pressure relief protection

Environment & utilities

Operating temperature range
−30 to +60 °C
Operating ambient humidity
20%–70% RH, non-condensing
Power supply
120 / 220 V AC, 50–60 Hz
CO₂ gas requirement
High-purity liquid CO₂ supply; recommended purity >99.9%

Cooling, control & filtration

Cooling system
Integrated closed-loop cooling; the internal chiller loop accelerates cooling and reduces CO₂ consumption
Control system
LCD touchscreen; semi-automated control with programmable timer operation and electronic sensor monitoring
Filtration system
Double T-Filter, Cat. No. 8785; two passes through 0.5 μm filters; >99.5% retention

Samples & chemical compatibility

Compatible sample types
Biological specimens, microscopy samples, hydrogels, aerogels, porous materials, polymers, and other structure-sensitive materials
Compatible solvents
CO₂-miscible solvents commonly used in sample preparation, including ethanol, acetone, methanol, and other laboratory solvents
Chemical compatibility warning
Strongly acidic, strongly alkaline, oxidizing, or highly toxic solvents require prior compatibility evaluation before use
Standard configuration & options

Standard configuration

  • High-pressure viewing window chamber
  • Integrated chamber illumination system
  • Integrated condensation and gas-liquid separation system
  • Non-mechanical CO₂ mixing and exchange system
  • Controlled slow filling and venting system
  • Precision vent flow meter
  • Standard stainless steel CO₂ supply line
  • Anti-static exhaust tubing
  • Replacement O-ring seal kit
  • Installation tool kit
  • User operation manual
  • Factory acceptance test report
  • Warranty and technical support

Optional configurations

  • Interchangeable sample holders
  • Custom sample chamber inserts
  • Magnetic stirring module
  • High-temperature / high-pressure chamber customization
  • Corrosion-resistant alloy chamber upgrade
  • Microscopy sample holder kit